System Analysis
As an example, let us consider the following measurement chain for a pressure measurement. It begins with the pressure, which a sensor converts into a voltage. Then the voltage is amplified and digitised.

Sensor
The sensor shows the following behaviour if no errors occur:
The measurement of the pressure is uncertain. The voltage has an offset of ±1 mV and an uncertainty in the slope of ±1% of the measured value. These are arbitrary values for the example. This gives us the formula for the sensor voltage:
The term (1 ± 0.01) has the value range [0.99 … 1.01]. This corresponds to [99% … 101%]. Mathematically, it maps the uncertainty of the slope of 1% of the measured value around 100% onto a range. The mapping does not yet give a distribution of the values within the range; it only marks the limits of the range.
The uncertainties are contained in the terms with “±”. To rearrange the formula so that the uncertainties are combined, we multiply out the bracket (1 ± 0.01):
The uncertainty is now combined in ΔUS. It has a constant term ±1mV. In addition there is a term that depends on the current pressure value P. For this example, let us insert P = 3 bar into the formula. This gives an uncertainty of the sensor voltage of
The deviation of the sensor voltage from the ideal characteristic at P = 3 bar lies in this interval. It is normally distributed, so it only extremely rarely reaches the limits of the interval and very probably lies close to the middle of the interval at 0mV.
Analogue signal processing
Suppose the pressure sensor can measure a pressure in the range P = [0 bar … 10 bar]. Then the value range of the ideal sensor voltage (without the uncertainty) is US,ideal = [0mV … 100mV]. Suppose the ADC has the reference voltage URef = 3V. To amplify the sensor signal optimally into this range, we need a gain of v = 30.
For this we use a non-inverting op-amp as an amplifier with
The resistors are not ideal; for a start we take ones in the 0.2% accuracy class with a temperature coefficient of TK=20ppm/°C. The pressure measurement is to be used at temperatures of T = [-25°C .. 75°C]. So ΔT = ±50°C around the reference temperature of 25°C applies. For the resistors this gives:
The op-amp is not ideal either; we use a type with an offset voltage of UOS = 5µV and a temperature coefficient of TK = 100nV/°C. The following applies:
We model the offset voltage at the non-inverting input of the op-amp, so it is amplified by the op-amp with v = 30.
ADC
For simplicity, we consider only the reference voltage of the 12-bit ADC to be erroneous. The 3V reference deviates initially by ±200ppm and has a temperature coefficient of TK = 10 ppm/°C.
Linearity, offset and quantisation effects of the ADC are neglected in this example.